EC-2000 Ultrasonic Thickness Gauge Meter
EC-2000, EC-2000-L and EC-2000-C is an instrument used to apply the principle of ultrasonic pulse reflection to measure the thickness (or speed of sound) of ultrasonic good conductor materials, including: measuring the thickness of metal materials such as steel, iron, copper, or aluminum, and measuring the thickness of non-metallic materials such as plastic, ceramic, rock, or glass.
EC-2000, EC-2000-C measuring range is 1~400mm, EC-2000-L is 1~300mm; The resolution of EC-2000 is up to 0.01mm, and the resolution of EC-2000-L and EC-2000-C is 0.1mm. When measuring, apply the coupling agent on the measured object, and then fit the probe to the daub part, and the instrument can measure the thickness of the measured object.
The EC-2000 uses a dot array display with a standard menu and data display interface for both average, minimum and maximum values. In addition, the instrument contains a large capacity memory. During the measurement process, the instrument can count and record the measurement results (the restart can be queried). EC-2000-L and EC-2000-C uses the segment code display screen, with a streamlined data display interface, the display effect is simple atmosphere, easy to operate and easy to use.
EC-2000, EC-2000-L and EC-2000-C standard probe line is about 1 meter long, which can be operated in a narrow space or assembly line.
|
Model |
EC-2000 |
EC-2000-L |
EC-2000-C |
|
Measuring principle |
Ultrasonic pulse reflection |
||
|
Frequency |
5MHz |
||
|
Measuring range |
1~400mm (steel) |
1~300mm (steel) |
1~400mm (steel) |
|
Measuring pattern |
Single/Continuous |
Continuous |
Single/Continuous |
|
Sound speed range |
1000~9999m/s |
||
|
Resolution ratio |
≤99.99mm: 0.01mm ≥100.0mm: 0.1mm |
0.1mm |
|
|
Accuracy |
1.00~9.99mm: ± 0.10mm 10.00~299.9mm: ±(1% of reading + 0.10mm) 300.0~400.0mm: ±(2% of reading + 0.10mm) Other range display values are for reference only |
1.00~9.99mm: ± 0.10mm 10.00~299.9mm: ±(1% of reading + 0.10mm) EC-2000-C: 300.0~400.0mm: ±(2% of reading + 0.10mm) Other range display values are for reference only |
|
|
Unit |
mm, inch |
||
|
Reaction time |
About 0.5 seconds for single measurement mode, about 0.3 seconds for continuous measurement mode |
||
|
Statistic |
100 groups of data |
60 groups of data |
|
|
User calibration |
4mm calibration |
||
|
Display |
128*128 Dot-matrix LCD screen, with a backlight |
Segment code screen, with a backlight |
|
|
Operational environment |
Temperature: -10~50℃, Humidity: ≤90%RH, no condensation dew |
||
|
Storage environment |
Temperature: -20~60℃, Humidity: ≤90%RH, no condensation dew |
||
|
Battery |
3 pcs AAA 1.5V alkaline batteries; USB 5V/2A (Type-C interface) |
||
|
Levels of protection |
IP40 |
||
|
Dimension |
135*72*27mm |
||
|
Case material |
ABS |
||
|
Weight |
About 127g (excluding battery) |
||